Lightspeed 1.X - Octane

Symptom

Anode or Cathode overcurrent errors when making X-ray. Anode overcurrent errors during tube warm up. Getting Diagnostic Errors 214321 and 214253 when running KV loop Diagnostic Thu May 18 11:38:42 1995 Suite: SAIC Host: OBC Proc: Diag Exec Error: 214321 File: technic_monitor.c 1.36 Method: technic_monitor Line: 1205 Function: Diagnostics XRAY GENERATION : MA Loop Test Name: Manual KV Unique ID: 48133 Iteration: 1 Exc. Level: Hard Time: 73779 Error Type: Hardware The maximum mA limit has been exceeded. Expected mA: 0mA Actual mA: Anode=121mA Cathode=121mA Maximum mA allowed: 5mA Thu May 18 11:35:28 1995 Suite: SAIC Host: OBC Proc: Diag Exec Error: 214253 File: isr1_handler.c 1.28 Method: obc_isr1_handler Line: 271 Function: Diagnostics XRAY GENERATION : KV Loop Test Name: XRay - Manual Unique ID: 48030 Iteration: 1 Exc. Level: Hard Time: 35142 Error Type: Hardware KV fault detected: Anode overcurrent. KV address: FFCFF3H Bit: D4

Error message

Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Method: No Method Line: 248 Function: XRAY GENERATION : KV Loop Scan: 18763/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 15:38:16:120 Log Series: 991 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3176T PERFORMIX LIGHTSPEED TUBE UNIT for QXI 1.x D3176T PERFORMIX LIGHTSPEED TUBE UNIT for QXI 1.x Part #: D3176T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX LIGHTSPEED TUBE UNIT FRU: Yes Status: Orderable Comments: If you have a CTI Need to order an D3132T. For LightSpeed 1X you must order D3176T. For LightSpeed 2X/3X/4X and HP60 please order D3186T *** SECOND OPINION NOTES *** Perform Rotor Control Diagnostics--If tube rotor fails to reach RUN mode in three ACCEL steps look for: High Drive output currents Low Drive output voltages and Low Drive output frequencies. This indicates a load problem with tube. With rotor running turn off Gantry 120V power at PDU. Listen for excessive noise and short deceleration time. A good rotor will coast for several minutes without power. Rotor issues (errors 183316, 183318, 183319, 260118111, 260118177) caused by tubes tend to be more prevalent on .5 second scans. If tube is installed on Jedi system, the Anode cable tends to break down (hard errors). Intermittent HV Spits -- Run KV Loop, 140KV, 30 sec, Anode, Cathode and both. Listen to tanks with stethoscope or screw driver, any noise indicates bad tank. Any spits? Remove HV cables from tanks fill with 2inch of oil. Spits = No, could be tube or cables. Spits = Yes, bad tank. Special Exception: No RMA will be auto-created for this part. Please ship back to warehouse via the normal process.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required