Pular para o conteúdo principal
GE Healthcare · MRDiscovery2267825

Erroneous emergency stop message generated by Auxiliary Board.

Symptom

Emergency stop condition is reported even though all subsystems are powered on Emergency stop condition is reported even though all subsystems are powered on..

Error message

2267825 The emergency stop button has been pressed or the emergency stop cable has been disconnected Fri Jan 16 12:13:40 2009 Error: 2267825Host: signa2 Process: NSPFile: SCP:FspMac.cpp Line: 320Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1The emergency stop button has been pressed or the emergency stop cable has been disconnected.->

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Component
5161071 XGD AUX PWB ENCL ASM Auxiliary Board in PGR Cabinet 5161071 XGD AUX PWB ENCL ASM.
Area
MR