Discovery NM 530c clinical scans for Thallium or Technetium have poor image quality Discovery NM 530c clinical…
Discovery NM 530c/570c
Symptom
Discovery NM 530c clinical scans for Thallium or Technetium have poor image quality Discovery NM 530c clinical scans for Thallium or Technetium have poor image quality. Investigation shows that the Thallium peak (70) and Technetium peak (140) were defined as 122 kev.
Applies to
- Models
- Discovery
- Product
- Discovery NM 530c/570c Discovery NM 530c/570c (CZT) DSV NM 530c DSV NM 570c
- Component
- *Software Software Software
- Area
- Nuclear Medicine