HEC detects PE and GC coolant flow high due to HEC switching to low power mode.

Symptom

HEC detects PE and GC coolant flow high Errors 2261713 and 2261724 are logged together. 2261713 ERROR: EM_HC_GRAD_COOLANT_FLOWRATE_HIGH_ALARM Heat Exchanger Cabinet (HEC) detected the gradient coil coolant flow is too high. Extended Error: Check the gradient coil coolant pressure sensor. 2261724 ERROR: EM_HC_XGD_COOLANT_FLOWRATE_HIGH_ALARM Heat Exchanger Cabinet (HEC) detected the power electronics coolant flow is too high. Extended Error: Check the power electronics coolant pressure sensor.

Error message

2261661 Cooling Cabinet switching to Low-Power mode after 15 minutes of inactivity 2261661 NSP SCP:CoolingHWO 923 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Probe-P Coil: HDNV Spectro Scan: 50252/13 Cooling Cabinet switching to Low-Power mode after 15 minutes of inactivity.

2261724 Heat Exchanger Cabinet HEC detected the power electronics coolant flow is too high. 2261724 NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Probe-P Coil: HDNV Spectro Scan: 50252/13 Heat Exchanger Cabinet (HEC) detected the power electronics coolant flow is too high. This alarm will not stop scanning-> Extended Error: Check the power electronics coolant pressure sensor.

2261713 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant flow is too high See Long Description 2261713 NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Probe-P Coil: HDNV Spectro Scan: 50252/13 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant flow is too high. This alarm will not stop scanning-> Extended Error: Check the gradient coil coo

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.