VRE Configuration Fails With System set to Non-English Language due to Software Bug.

Symptom

VRE Configuration Fails With System set to Non-English Language See Long Description iTrak # 13230392 15.0 m3 software: 15.0M3 software was being installed as an upgrade from 14.0M5a. As per instructions, operating system and application software had been loaded and RestoreInfo had been performed. The system had been configured to run in French with the 14x software so this setting had been imported when the RestoreInfo was done. VRE Configuration failed with no error messages, just a dialog box with an OK button and "VRE Configuration Failed." iTrak # 13224435 14.0 M5 software: WHEN THE USER INTERFACE LANGUAGE IS SET TO FRENCH WE ARE NOT ALLOWED TO CONFIGURE THE VRE or ICN

Error message

VRE Configuration Failed See Long Description VRE Configuration Failed VRE Configuration failed with no error messages, just a dialog box with an OK button and "VRE Configuration Failed."

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only