GE Healthcare · MROvation
Part Numbers for Ferrite Core Kits difficult to Find.
Symptom
Part Numbers for Ferrite Core Kits difficult to Find Ferrite cores are sometimes used to troubleshoot noise issues which cause low SNR on Ovation and Profile. The cores are placed on cables at the PenPanel to dampen electrical noise riding on the cables shield. Since these cores are made of iron, it is not recommended to use near the magnet.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Component
- 2138402 CORE CLAMP ASSY Ferrite cores Contains 5 ferrite cores of different sizes. Order more than one bag. ICV/Transfer Price: 73.00 USD /Each U0028MC MISC ELECTRONIC COMPONENT Ferrite Core Contains 10 ferrite cores all of the same size. Order more than one bag. ICV/Transfer Price: 5.60 USD /Each
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100060 Unable To Film.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.
- Having Difficulties Setting Up Gradcal Scan From Dir 5112585 - Signa Ovation With Excite Technology
- CWA ID101091. Some or All Images in the Last Scan Did Not Reconstruct.
- HFO Cradle Will Not Move Due To Loss Of Home Flag Position.