Unable to Complete Head or Body UPM Reflected Power Calibration due to Error, UPM Calibration failed due to failure to c…
Unable to Complete Head or Body UPM Reflected Power Calibration due to Error, UPM Calibration failed due to failure to converge on a calibration point due to the FE Left the Dummy Load Connected to th…
Symptom
Unable to Complete Head or Body UPM Reflected Power Calibration with UPM Calibration failed due to failure to converge on a calibration point See Long Description Unable to complete Head or Body UPM Reflected Power Calibration; results wont converge. The following error message is logged: (Note: the actual UPM number is substituted for the wildcard "%d") 2257265 "UPM %d Calibration failed due to failure to converge on a calibration point."
Error message
2257265 UPM 2 Calibration failed due to failure to converge on a calibration point. Error: 2257265 Process: NSP File: SCP:Upm_Cal_Test Line: 1052 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: cal Coil: HEAD Scan: 50280/1 UPM 2 Calibration failed due to failure to converge on a calibration point. Verify that power is applied on the proper channel on the UPM.->
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.