123456 DLX not communicating check VDP power
Symptom
DLX not communicating check VDP power Had fe remove all coax cables from Sun transceiver tee and place 50 ohm terminator in place and run test net-aui, the test passed both internal and external test FE now moving cables and transceiver around to find which one fails found 50 ohm self terminator on titan kernal bad FE had critical hardware error did dlxtest, vdp conifiguration found disk 0 not present. see long description for more details Remove one of the coax tee with the coaxs still attached and measure the center pin to the shield this shoul measure 25 ohms Had fe remove all coax cables from sun transceiver tee and place 50 ohm terminator in place and run test net-aui the test passed both internal and external test fe now moving cables and transcever around to find which one fails found 50 ohm self terminator on titan kernal bad fe had critiical hardware error did dlxtest vdp conifiguration found disk 0 not present fe checking the disk connections since he had pulled out the the vdp cpu card prior to calling olc see xr-solution 019930
Error message
"Check DLX Cabinet Power" error message displayed on Flat screen "Check DLX Cabinet Power" error message displayed on Flat screen. Communication problem between SUN and VDP at DLX application start up (i.e after typing DLX at DLX login prompt).
Applies to
- Models
- Advantx
- Product
- DLX 3 DLX 3 digital system DLX 3 digital system
- Software
- DLX C19.07 DGW Sw. Jumpstart Software JumpStart Apps Apps Version Rev Kit Floppy CD Solaris Floppies ----------- ----------- ---------- ---------- --------- --------- --------- 4.02 3.0 2286500 2267136 2286489 2211863 N/A
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
- Intermittent Gantry Lockup usually occurred when bagging the Detector or Removing the patient from the table.
- Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
- Omega, Can not move the table top in the lateral direction.
- The Advantx Console is Blank on boot of the system.
- Exabyte won't archive or eject