Pular para o conteúdo principal
GE Healthcare · MROvation2225932

CWA ID 123073 ECG and Cardiac Gating have no Signal.

Symptom

ECG and Cardiac Gating have no Signal Cannot change from peripheral gating to ECG gating (any lead), no signal. ECG and Cardiac Gating have no Signal

Error message

2225932 SPU Cardiac Process not responding to IMP commands. Error: 2225932 Host: SMR2-OC0 Process: NSP File: HPC_PORT_ID:spu_impif Line: 7 Internal Software Error. A reset of TPS may be necessary. SPU Cardiac Process not responding to IMP commands. Command = 5.

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO2 Ovation2 with Octane2
Area
MR

Full cause & step-by-step resolution