Application software will not boot Due to SCP Board.

Symptom

HP Computer Will Not Boot Applications Software

Error message

2252112 Board SCP is in invalid power up state 2 See Long Description Verify ENV parameters and retry power up. Extended Error: The board may be in one of the following states: 0 - No response, 1 - OS Loaded, 2 - PPC Bug, 3 - PPC Diag, 4 - OS not loaded If the error message states that the board is in state 4, then the OS load was stopped. This may happen if characters are detected on the serial cable. In order to eliminate this problem, disconnect the serial cable from AGP/COM1 Connector or SCP/Console Connector and reboot the board. If the error message states that the board is in state 2 or 3, then run the Motorola Board Level Diagnostics to verify the board is OK. If the BLD passes but this message appears during TPS reset then it is highly probable that the boards are not being reset by STIF connection. You can verify if the boards are not being reset by STIF connection during TPS reset either by looking at the LEDs on front of the boards or by opening MGD Terminal window.

2223849 SCP EPROM power-up boot failed (See attachment for extended error message) 2223849 rmny NSP TardisReset.cc 1841 Exception Class: Unknown Function: Host: NSP SCP EPROM power-up boot failed. Communica

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OVATION5 Applies to Ovation5 (Excite) only. Ovation5 is 12.x platform software, not OvationHD.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.