GE Healthcare · CT18310550005

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Stopped Scan message DURING Scan data acquisition. Long Description for Feedback Only. Scanner Hardware Stopped Scan message DURING Scan data acquisition.

Error message

Error: 183105 KV fault detected: Anode overvoltage. Anode kV > 82kV. See LD for error example Thu Sep 16 11:15:54 2004 Error: 183105 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 210 Function: XRAY GENERATION : KV Loop Scan: 50005/9/13 Type: Axial Exception Level: Pri/Most Severe Time: 11:15:55:750 Log Series: 373 KV fault detected: Anode overvoltage. Anode kV > 82kV. Fault latch address = ffcff3 hex, Bit = D0

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2203602 (replaces 2161307) HEMIT HV TANK used on CT/I (performix), HiSpeed QXi, LS1.X LS2.X, LS 3.X, LS4.X See Long Description for details Part #: 2203602 Part Type: GP REPAIRABLES Part Weight: 55.12 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HEMIT HV TANK SPARE PART FRU FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The HV adapter/Spacer part number 2171866 Kit that is included with the HV tanks But it is only used on older Highspeed Gantry due to the smaller bolt size. Refer to MI&CT ESERVICE NOTE SN07091114a for additional details. (212001261 21-Jan-2009) *** 2ND OPINION NOTES *** Take an exposure and listen to tank. If you hear an arcing sound most likely bad tank. Rotate Gantry so tank is upside down and take exposure. Listen for arcing. Turn off lights in room and check for a flash from the tank. Remove tank, set on floor on top of the lifting bracket. Make room quiet. Rock tank front to back and side to side, listen carefully. Any sounds, indicates a failure. Check tank for oil leaks Old FRU Compatible No Comments ct/i helio rspl Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable Yes Use on Hands No Use Demand History Yes Ultimate Prime 2203602 New FRU Backward Compatible Yes Old FRU Usable No Change Type Direct ECO Number 9999
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.