GE Healthcare · CT178588
Lightspeed 1.X - Octane
Symptom
No flashing eights on gantry display and Scanning hardware reset failed
Error message
Error: 178588 Failed to download charaterization file 4-fw/config/Ro tor.cfg Tue Oct 11 01:18:55 2005 Error: 178588 Host: ct02 Process: fwmgr File: Characterization.c Line: 280 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to download charaterization file 4-fw/config/Ro tor.cfg Wed Apr 1 13:44:43 2009 Error: 178588 Host: sea_rad_ct1 Process: fwmgr File: Characterization.c Line: 281 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to download charaterization file 13-fw/char/CalPts.char
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-321300G1 SCOM FOR USE WITH ZEUS RP See Long Discription for Details 46-321300G1 SCOM FOR USE WITH ZEUS RP Part #: 46-321300G1 Part Type: GP REPAIRABLES Part Weight: 2.27 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: SCOM FOR USE WITH ZEUS RP FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution