Unable to Scan due to Error Code 2267906 due to One of the Leak Sensor Strips in Base of PGR Cabinet was Disconnected.

Symptom

Unable to Scan With Error 2267906 The Power Cabinet AUX board is reporting a leak sensor cable-off fault. See Long Description Unable to scan due to error shown below, which is logged in the message log. 2267906 LX-BAY11B NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Power Cabinet AUX board is reporting a leak sensor cable-off fault. TPS Reset and reattempt scan.

Error message

2267906 The Power Cabinet AUX board is reporting a leak sensor cable-off fault. See Long Description 2267906 LX-BAY11B NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Power Cabinet AUX board is reporting a leak sensor cable-off fault. TPS Reset and reattempt scan.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR