Pular para o conteúdo principal
GE Healthcare · CT230004020230006515230003513230003002230003011200352906

Lightspeed 1.X - Octane

Symptom

Recon self test failure on boot up. Recon self test failure on boot up

Error message

Error: 230004020 Unable to load IG unit 0 with file See long description for full sample error. Tue Jul 22 00:58:13 2008 Error: 230004020 Host: ice Process: igproxy File: igproxy events.c Line: 1602 Unable to load IG unit 0 with file l C2 H l C2 c l C2 e l C2 g l C2 i l C2 k l C2 m l C2 o l C2 q l C2 s

Error: 230006515 WARNING: Reconmgr receives a queue notify while it is not in IDLE state See long description for full sample error. Tue Jul 22 09:27:59 2008 Error: 230006515 Host: ct01 sbc Process: reconmgr File: rm recon events.c Line: 851 Function: Image Generation : SBC Processing Job Type: Not defined Scan: 0/0/0 Image: 0/0/0 Recon Seq Id: 0 WARNING: Reconmgr receives a queue notify while it is not in IDLE state.

Error: 230003513 Error -1 loading IG 0 See long description for full sample error. Tue Jul 22 00:58:13 2008 Error: 230003513 Host: ct01 sbc Process: igctrl File: igproxy if.c Line: 755 Function: Image Generation : SBC Processing Job Type: Not defined Scan: 0/0/0 Image: 0/0/0 Recon Seq Id: 0 Error -1 loading IG 0

Error: 230003002 IGCTRL failed in the process: load all igs() with status: -1 See long description for full sample error. Tue Jul 22 00:58:13 2008 Error: 230

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues