Pular para o conteúdo principal
GE Healthcare · MROvation22410792240410

PC Will Not Boot

PC Will Not Boot.

Symptom

PC Will Not Boot

Error message

2241079 Workspace Interface Module received a bad opcode from Workspace PC over serial link. See Long Description Error: 2241079 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_wim Line: 171 Workspace Interface Module received a bad opcode from Workspace PC over serial link. Opcode = 0x0. First, run diagnostics to test Workspace Interface Module to Workspace PC communication. If error persists, check cable connections, power cycle Workspace Interface Module and reboot Workspace PC.

2240410 - Host to Workspace PC Process Error. See Long Description Error: 2240410 Host: lx-mr Process: hop_server File: cpb_proto.c Line: 991 Host to Workspace PC Process Error. The Host received a unknown opcode from the Workspace PC. Verify Host to Workspace PC communication cables. Shutdown and boot host and Workspace PC.

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
Component
2284297 - MR OPERATOR WORKSPACE PC COMPUTER 2284297 - MR OPERATOR WORKSPACE PC COMPUTER 2284297 - MR OPERATOR WORKSPACE PC COMPUTER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required