Diffusion Scans Abort with XPS Gradient Distortion Faults and XPS Overvoltage on Discovery MR750 Systems due to Site Fai…

Diffusion Scans Abort with XPS Gradient Distortion Faults and XPS Overvoltage on Discovery MR750 Systems due to Site Fails Preinstallation Power Requirement.

Symptom

Diffusion Scans Abort with XPS Gradient Distortion Faults and XPS Overvoltage or XPS Overcurrent Faults on Discovery MR750 Systems See Long Description EPI diffusion-weighted scans frequently abort with one or more of the following error codes logged: 2267888 The %b XPS is reporting a Low Voltage I-Bus Over-current Fault.-> 2267893 EM_XGD_PS_VBUS_IN_FAULT The %b XPS-SB is reporting an internal VBus overvoltage warning 2267857 A gradient distortion fault has occurred on the %b.->

Error message

2267888 The %b XPS is reporting a Low Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persist replace the failing XPS. Note: In the actual error message seen on the scanner, "%b" will be replaced with either the X, Y, or Z axis.

2267857 A gradient distortion fault has occurred on the %b. See Long Description 2267857 "A gradient distortion fault has occurred on the %b.->" Extended Error: Ensure all communication channels are functioning properly and that the gradient coil is connected and functioning properly.

2267893 The %b XPS-SB is reporting an internal VBus overvoltage warning See Long Description 2267893 EM_XGD_PS_VBUS_IN_FAULT The %b XPS-SB is reporting an internal VBus overvoltage warning

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required