GE Healthcare · X-ray VascularAdvantx100007

The Applicaiton will not boot.

Symptom

DLX: The Applicaiton will not boot. After DLX log in the DLX application begins to load, and the system resets. The DLX application will not boot. After DLX log in the DLX application begins to load, and the system resets back to the log in prompt. The site is unable to launch the DLX application.

Error message

DLX: Error 100007 A Process or GEIM software crashed (Application restarted) is logged in the error logs. # filename: code.100007 # generate: from_interleaf version V1.2 # mnemonic: USUP_Process_Error # category: Crash_hang # severity: Severe # label : "A Process or GEIM software crashed (Application restarted)" # CAUSE : unknown # ACTIONS : See TSG003C. Perform GEIMEXER of DLXTEST. # COMMENTS: Probing mechanism failed

Applies to

Models
Advantx
Product
DLX DLX DLX
Software
DLX C22.09 DGW Sw. Jumpstart Software JumpStart Apps Apps Version Rev Kit Floppy CD Solaris Floppies ----------- ----------- ---------- ---------- --------- --------- --------- 5.02 4.0 2276196 2281552 2276194 2211863 N/A
Area
X-ray Vascular