When booting the dlx the System is reporting
Symptom
DLX: When booting the dlx the System is reporting: File system full at login. When booting the dlx the System is reporting: File system full at login.The console message is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " See Sn 4559 for leap day issue When booting the dlx the System is reporting: File system full at login. The console message is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " If the DLX is rebooted when this error message first appears, the system will fail subsequent boots and reset. In order to recover, the system must be booted in single user mode. To do this use the "stop + a" keys to halt the system at the OK> prompt boot the system to single user mode using the "boot -s" command. when the system halts at the log in or proceed with normal start up question enter the root operator password. After the password is entered the system will go to the # prompt. See Sn 4559 for leap day issue
Error message
DLX: File System Full error message. No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full DLX: File System Full error message. After booting the DLX when the system gets to the login prompt it reports "File System Full". Do not reboot, if possible, the system if this error is present. The message displayed on the DLX monitor is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " DLX: File System Full error message. After booting the DLX when the system gets to the login prompt it reports "File System Full". Do not reboot, if possible, the system if this error is present. The message displayed on the DLX monitor is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full "
Applies to
- Models
- Advantx
- Product
- DLX 3 DLX 3 digital system DLX 3 digital system
- Software
- DLX C18.17 DGW Sw. Jumpstart Software JumpStart Apps Apps Version Rev Kit Floppy CD Solaris Floppies ----------- ----------- ---------- ---------- --------- --------- --------- 3.03 2.3 2254104 2241330 2252563 2211863 2252564
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
- Intermittent Gantry Lockup usually occurred when bagging the Detector or Removing the patient from the table.
- Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
- Omega, Can not move the table top in the lateral direction.
- The Advantx Console is Blank on boot of the system.
- Exabyte won't archive or eject