Pular para o conteúdo principal
GE Healthcare · X-ray VascularAdvantx

When booting the dlx the System is reporting

Symptom

DLX: When booting the dlx the System is reporting: File system full at login. When booting the dlx the System is reporting: File system full at login.The console message is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " See Sn 4559 for leap day issue When booting the dlx the System is reporting: File system full at login. The console message is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " If the DLX is rebooted when this error message first appears, the system will fail subsequent boots and reset. In order to recover, the system must be booted in single user mode. To do this use the "stop + a" keys to halt the system at the OK> prompt boot the system to single user mode using the "boot -s" command. when the system halts at the log in or proceed with normal start up question enter the root operator password. After the password is entered the system will go to the # prompt. See Sn 4559 for leap day issue

Error message

DLX: File System Full error message. No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full DLX: File System Full error message. After booting the DLX when the system gets to the login prompt it reports "File System Full". Do not reboot, if possible, the system if this error is present. The message displayed on the DLX monitor is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full " DLX: File System Full error message. After booting the DLX when the system gets to the login prompt it reports "File System Full". Do not reboot, if possible, the system if this error is present. The message displayed on the DLX monitor is "No UTMPX entry you must exec login from lowest level shell NOTICE realloccg /: file system full "

Applies to

Models
Advantx
Product
DLX 3 DLX 3 digital system DLX 3 digital system
Software
DLX C18.17 DGW Sw. Jumpstart Software JumpStart Apps Apps Version Rev Kit Floppy CD Solaris Floppies ----------- ----------- ---------- ---------- --------- --------- --------- 3.03 2.3 2254104 2241330 2252563 2211863 2252564
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues