Alarm (high) — pump-stopping faultWater Cooling System (Oxford 2195 Cooler Cabinet)

POLPump Overload alarm on the Oxford 2195 cooler cabinet

Philips CT error code, reported by the Oxford 2195 Cooler Cabinet (MRI gradient/primary water cooling).

What POL means

On Eclipse 1.5T and Polaris and Apollo CT systems, status code POL indicates: Pump Overload alarm on the Oxford 2195 cooler cabinet. The code originates in the Water Cooling System (Oxford 2195 Cooler Cabinet) section of the Oxford 2195 Cooler Cabinet (MRI gradient/primary water cooling). Severity level alarm-high (Alarm (high) — pump-stopping fault).

Reported action: Pump Overload indicates the circulation pump has tripped on overload. Acknowledge outstanding alarms from the controller keypad by pressing 9 followed by Return, then command the pump to stop and start again before returning the cooler to service. This document does not detail further pump-specific diagnostic steps..

Shown as an alarm-status abbreviation on the Picker 3020 controller of the Oxford 2195 Cooler Cabinet, documented for the Philips/Picker Eclipse 1.5T MRI (and shared, per the same table, by Polaris and Apollo systems). Not confirmed by this document for other cooler models.

Full troubleshooting for POL

The complete diagnosis, cause tree, and step-by-step remedy for the Water Cooling System (Oxford 2195 Cooler Cabinet) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Water Cooling System (Oxford 2195 Cooler Cabinet) codes

Discuss POL with other field engineers in the FSELIB forum

Source: Philips/Picker Eclipse 1.5T MRI service documentation — Theory and Troubleshooting of Oxford 2195 Cooler Cabinet (Diagnostic Tools and Information, Maintenance). Reference information re-authored by FSELIB — not a copy of the OEM manual.